3.2 Remote control commands

795 KFT Titrino

88

.T.Si

SiloEmpty: Silo empty, i.e. the last line has been removed from

the silo memory.

.T.M

MeasList: Entry in the measuring point list.

.T.EP

EPList: Entry into EP list

.T.RC

Results have been recalculated.

Messages from node .C, Comport:

.C.B1

COM1: A report is outputted on COM1. During this time, COM2

will be blocked. COM2 is generally blocked, if COM1 is busy.

.C.R1

COM1 is ready again. (Comes also when you <QUIT> an error.)

.C.B2, .R2 Identical for COM2.

Messages for changings in the I/O lines. If the changings are made simulta-

neously, there is 1 message. Pulses receive 2 messages: one message each

for line active and inactive.

.I

Input: Change of an input line.

.O

Output: Change of an output line (except 7, pin 2, for recorder

pulses).

If a change occurs that requires a message, the Titrino sends space

(ASCII 32) and ! as an introducer. This is followed by the name of the device

(see 3.2.2.47). Special ASCII characters in the device name are ignored. If no

device name has been entered, only ! is sent. Finally the Titrino sends the

information which node has triggered the message.

Example: !John".T.Si": The message was triggered from instrument "John",

node .T.Si

3.2.2.93. Setup.Graphics.Grid ON, OFF

Setup.Graphics.Frame ON, OFF

Setup.Graphics.Scale Full, Auto

Setup.Graphics.Recorder.Right 0.2...0.5...1.00

Setup.Graphics.Recorder.Feed 0.01...0.05...1.00

Change in the appearance and the format of the curve for the output. The set-

tings are valid for both Titrino COM ports.

.Grid:

On/off switching of grid over curve.

.Frame:

On/off switching of frame surrounding the curve. If grid and frame

are switched off, the curve is printed faster as the print head does

not have to move to the end of the paper.

.Scale:

Type of scaling of the measured value axis: Full means that the

scale runs from the smallest up to the greatest measured point.

With auto, the smallest measured value is taken and the next

smaller tick defines the beginning of the scale; the next greater

tick to the greatest measured value is the end of the scale.

.Right:

Relative specification of the width of the output medium (e.g. pa-

per width) for the length of the measured value axis. 1 means the

measured value axis is plotted over the entire width of the paper

(largest possible width). In extreme cases, the writing of the right

tick may lie outside.