iEC 60601 T

EsT

l

EvEl

C

omplianCE

l

EvEl

E

lECTromagnETiC

E

nvironmEnT

- g

uidanCE

Electrostatic

Discharge (ESD)

IEC 61000-4-2

±(2,4,6) kV contact

±(2,4,8) kV air

±(2,4,6) kV

contact

±(2,4,8) kV air

Floors should be wood,

concrete or ceramic tile. If floors

are covered with synthetic

material, the relative humidity

should be at least 30%.

Electrical Fast

Transient/Burst

IEC 61000-4-4

±2 kV for power

supply lines

±1 kV for input-

output lines

±2 kV for power

supply lines

±1 kV for input/

output lines

Mains power quality should be

that of a typical commercial or

hospital environment.

Surge

IEC 61000-4-5

±1 kV differential

mode

±2 kV common

mode

±1 kV

differential

mode

±2 kV for

common mode

Mains power quality should be

that of a typical commercial or

hospital environment.

Voltage

dips, short

interruptions

and voltage

variations on

power supply

input lines

IEC 61000-4-11

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in U

T

)

for 5 sec

<5% U

T

(>95% dip in

U

T

) for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in

U

T

) for 5 sec

Mains power quality should

be that of a typical commercial

or hospital environment. If the

user of the goLITE BLU requires

continued operation during

power mains interruptions,

it is recommended that the

goLITE BLU be powered from an

uninterruptible power supply or

a battery.

Power

frequency

(50/60 Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic

fields should be at levels

characteristic of a typical

location in a typical commercial

or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.

27

i

mmuniTy

T

EsT

iEC 60601

T

EsT

l

EvEl

C

omplianCE

l

EvEl

E

lECTromagnETiC

E

nvironmEnT

- g

uidanCE

Conducted

RF

IEC 61000-

4-6

Radiated

RF

IEC 61000-

4-3

3 Vrms

150 kHz to

80 MHz

3 V/m

80 MHz to

2.5 GHz

3 V

3 V/m

Portable and mobile RF communications

equipment should be used no closer to any

part of the goLITE BLU, including cables,

than the recommended separation distance

calculated from the equation applicable to the

frequency of the transmitter.

Recommended separation distance:

d = 1.2

d = 1.2

d = 2.3

Where P is the maximum output power rating

of the transmitter in watts (W) according to

the transmitter manufacturer and d is the

recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site survey

a, should be less than the compliance level in

each frequency range b.

Interference may occur in the vicinity of

equipment marked with the following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects, and people.

a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and

land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted

theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,

an electromagnetic site survey should be considered. If the measured field strength in the location

in which the goLITE BLU is used exceeds the applicable RF compliance level above, the goLITE BLU

should be observed to verify normal operation. If abnormal performance is observed, additional

measures may be necessary, such as re-orienting or relocating the goLITE BLU.

b: Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.

GUIDANCE AND MANUFACTURER’S DECLARATION —

ELECTROMAGNETIC IMMUNITY: The goLITE BLU is intended for use in

the electromagnetic environment specified below. The customer or the

user of the goLITE BLU should make sure it is used in such an environment.