GUIDANCE AND MANUFACTURER’S DECLARATION —

ELECTROMAGNETIC IMMUNITY: The goLITE BLU is intended for use in

the electromagnetic environment specified below. The customer or the

user of the goLITE BLU should make sure it is used in such an environment.

i

mmuniTy

T

EsT

iEC 60601 T

EsT

l

EvEl

C

omplianCE

l

EvEl

E

lECTromagnETiC

E

nvironmEnT

- g

uidanCE

Electrostatic

Discharge (ESD)

IEC 61000-4-2

±(2,4,6) kV contact

±(2,4,8) kV air

±(2,4,6) kV

contact

±(2,4,8) kV air

Floors should be wood,

concrete or ceramic tile. If floors

are covered with synthetic

material, the relative humidity

should be at least 30%.

Electrical Fast

Transient/Burst

IEC 61000-4-4

±2 kV for power

supply lines

±1 kV for input-

output lines

±2 kV for power

supply lines

±1 kV for input/

output lines

Mains power quality should be

that of a typical commercial or

hospital environment.

Surge

IEC 61000-4-5

±1 kV differential

mode

±2 kV common

mode

±1 kV

differential

mode

±2 kV for

common mode

Mains power quality should be

that of a typical commercial or

hospital environment.

Voltage

dips, short

interruptions

and voltage

variations on

power supply

input lines

IEC 61000-4-11

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in U

T

)

for 5 sec

<5% U

T

(>95% dip in

U

T

) for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in

U

T

) for 5 sec

Mains power quality should

be that of a typical commercial

or hospital environment. If the

user of the goLITE BLU requires

continued operation during

power mains interruptions,

it is recommended that the

goLITE BLU be powered from an

uninterruptible power supply or

a battery.

Power

frequency

(50/60 Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic

fields should be at levels

characteristic of a typical

location in a typical commercial

or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.

4222.002.7504.1.indd 26

29-05-09 15:56