5.12 Properties of the RS232 Interface
774 Oven Sample Processor as Sender:
Time
RxD
TxD
Data output
Data input
LF
LF
LF
Data output
disabled
Data output
enabled
Request
1.Line
Response
XON
XOFF
external device
Sample Changer
Changer transmission can be stopped by external instruments with XOFF.
After XOFF is received the changer completes sending the line already
started. If data output is disabled for more than 3 s by XOFF, E43 appears
in the display.
Hardware-Handshake, HWs
774 Oven Sample Processor as Receiver:
Time
DTR
LF
RxD
RxD
external device
Sample Changer
DTR
774 Oven Sample Processor as Sender:
Time
RTS
RTS
LF
TxD
TxD
CTS
CTS
external device
Sample Changer
The data flow can be interrupted by deactivating the CTS line.