3.2 Remote control commands

798 MPT Titrino

119

Assembly.Bur.ModeDis.VStop 0.0001...9999, OFF

Assembly.Bur.ModeDis.AutoFill ON, OFF

Dispensing mode with parameters. The dispensing mode can only be started

and stopped via the RS Control. During a running dosification, no method can

be started at the Titrino.

.Selected: Dispensing of volume increments or during a preset time.

.Volume, .Time: Size of the volume increments or entry of time.

.VStop:

Limit volume for the dispensing.

.AutoFill: ON means automatic filling after every dispensing.

3.2.2.112. Assembly.Meas.Status ON, OFF

Assembly.Meas.MeasInput

1, 2, Diff., Ipol, Upol, Temp

Assembly.Meas.Ipol ±127...1...+127

Assembly.Meas.Upol ±1270...400...+1270

Measurement in assembly. The measuring function can only be started via RS

Control. When the measuring function is switched on, no method can be

started at the Titrino.

.Input:

Selection of the potentiometric measuring input 1, 2, diff., polar-

ized electrodes or temperature.

.Ipol:

Polarization current in µA.

.Upol:

Polarization potential in mV, entry in steps of 10 mV.

3.2.2.113. Assembly.Outputs.AutoEOD ON, OFF

Assembly.Outputs.SetLines $G

Assembly.Outputs.SetLines.L0

active, inactive, pulse, OFF

up to .L13

Assembly.Outputs.ResetLines $G

Setting the I/O output lines.

.AutoEOD: The automatic output of the EOD (End of Determination) at the

end of the determination can be switched off. Thus, for example,

in conjunction with a Titrino several determinations can be per-

formed in the same beaker. Before AutoEOD is switched on, line 3

must be set to "OFF".

.SetLines: With $G, all lines are set.

.SetLines.LX: Set the line LX. "active" means setting of a static signal, "inac-

tive" means resetting of the signal, "pulse" means output of a

pulse of app. 150 ms, "OFF" means the line is not operated, see

also page 167.

Warnings:
• If you have "AutoEOD" to "ON", an active line 3 is set to "inac-

tive" by the EOD pulse.

• L6 is the line of the activate pulse. An active line 6 is set to "in-

active" by the activate pulse.

• L5 is the error line. It is continuously controlled by the Titrino

program and can therefore not be set freely.

Line assignments in Titrino program:

L0

Ready, inactive state

L1 Conditioning

OK

L2

Titration in progress

L3

EOD (End Of Determination)