3.2 Remote control commands

798 MPT Titrino

105

3.2.2.43. Mode.Parameter.Calibration.SignalDrift 0.5...2...999, OFF

Mode.Parameter.Calibration.EquTime 0...110...9999, OFF

Parameters for CAL: Criteria for measured value acquisition. Measured value

drift in mV/min, equilibration time in s. OFF means that the corresponding cri-

terion is switched off. If both criterions are on OFF, the measured value is ac-

quired immediately.

If the equilibration time has never been edited, it is automatically calculated by

the instrument to match the drift, see page 19. After it has been edited once, it

remains in force with the set value.

3.2.2.44. Mode.Parameter.Calibration.ElectrodeId

up to 8 ASCII char

Parameters for CAL: Electrode identification. It is classified under calibration

data, see 3.2.2.93.

3.2.2.45. Mode.Parameter.Calibration.SmplChanger ON, OFF

Parameters for CAL: Calibration at Titrino.

With "ON", there are no hold points in the calibration sequence for entries, the

first buffer is measured directly.

3.2.2.46. Mode.Parameter.Calibration.ActPulse first,

all, OFF

Parameters for CAL: Output of a pulse on the I/O line "Activate", see page

167.

3.2.2.47. Mode.Parameter.Sequence.X.Select

method, pause, L4 output,

L6 output, info, stirrer, OFF

Parameters for TIP: Selection of an element for step X (X = 1...30). For the

parameters of the elements see 3.2.2.48.

3.2.2.48. Mode.Parameter.Sequence.X.Method Method

name

Mode.Parameter.Sequence.X.Pause

0...999 999, INF

Mode.Parameter.Sequence.X.L4Output

active, inactive,

pulse, OFF

Mode.Parameter.Sequence.X.L6Output

as for L4

Mode.Parameter.Sequence.X.Info

up to 16 ASCII characters

Mode.Parameter.Sequence.X.Stirrer

ON, OFF

Parameters for TIP: Parameters of the elements of TIP.

.Method: Method name of a method available in the user memory or on the

card. Up to 8 ASCII characters.

.Pause:

Pause time in s. INF means infinite. Continue the sequence with

&m $G.

.L4 Output: Warning: A pulse triggered by the limit value monitoring at L4 (pin

3) in a submethod sets an output set to active in TIP to inactive.

.L6 Output: Warning: An activate pulse at L6 output (pin 1) in a submethod

sets an output set to active in TIP to inactive.

.Info:

Entry of a message which is written into the display. The se-

quence remains in the display with the corresponding message.

Continue with &m $G.

Advertising