6-3. Etalonnage lorsque la fonction HOLD est activée

IM 12D7C22-F-H

Calibration 6-3

µ

S / c m

HOLD

µ

S / c m

ENT

HOLD

µ

S / c m

ENT

HOLD

>

Select the flashing digit with the > key.
Increase its value by pressing the key

>

Set the value
using the >, , ENT key.

µ

S / c m

NO

YES

HOLD

YES

µ

S / c m

NO

YES

HOLD

Put the sensor in standard
solution. Press YES.

When the correct value is displayed,
press ENT to enter the change.

After briefing displaying WAIT,
the CAL.END message appears.

The calibration is now complete. Put the
sensor back in the process and press YES.

µ

S / c m

NO

YES

HOLD

HOLD will be displayed. Press NO to turn off
HOLD and return to the measuring mode.

µ

S / c m

NO

YES

HOLD

µ

S / c m

NO

YES

NO

HOLD

NO

MODE

YES

ENT

YOKOGAWA

MODE

CONTACTS

S1

S2

S3

FAIL/S4

µ

S / c m

Press the MODE key.

The legend CAL.1/CAL.2

appears, and the YES/NO

key prompt flags flash.

If the SETP legend

appears, press NO first.

MODE

MEASURE

DISPLAY 1

HOLD

DISPLAY 2

CAL 2

CAL 1

Note: See section 4-5for full

display acces function

Appuyer sur MODE
CAL1/CAL2 apparaît YES/NO
clignote. Se SETP s’affiche,
appuyer d’abord sur NO

Plonger le capteur dans la
solution standard. Appuyer
sur YES.

Régler la valeur à l’aide des touches >, ^ et ENT

Sélectionner le digit clignotant avec > et augmenter
sa valeur à l’aide de ^.

Valider avec ENT

WAIT s’affiche brievement CAL.END apparaît

L’étalonnage est terminé. Replonger le capteur
dans le proceé. Appuyer sur YES pour repasser en
mode mesure.

HOLD s’affiche, appuyer sur NO pour revenir au
mode mesure.

Note: voir section 4-5.