6-2. Procédure d’étalonnage

6-2 Etalonnage

ENT

M

Ω .

c m

ENT

M

Ω .

c m

>

Select the flashing digit with the > key.
Increase its value by pressing the key

>

Set the value
using the >, , ENT key.

NO

YES

M

Ω .

c m

NO

YES

M

Ω .

c m

Put the sensor in standard
solution. Press YES.

When the correct value is displayed,
press ENT to enter the change.

After briefing displaying WAIT,
the CAL.END message appears.

The calibration is now complete. Put the
sensor back in the process and press
YES to return to the measuring mode.

ENT

M

Ω .

c m

M

Ω .

c m

YES

NO

YES

M

Ω .

c m

NO

NO

MODE

YES

ENT

YOKOGAWA

MODE

CONTACTS

S1

S2

S3

FAIL/S4

M

Ω .

c m

CAL 2

CAL 1

MEASURE

DISPLAY 1

HOLD

DISPLAY 2

Press the MODE key.

The legend CAL.1/CAL.2

appears, and the YES/NO

key prompt flags flash.

If the SETP legend

appears, press NO first.

MODE

Note: See section 4-5 for full

display access details

Appuyer sur MODE
CAL1/CAL2 apparaît YES/NO
clignote. Se SETP s’affiche,
appuyer d’abord sur NO

Plonger le capteur dans la
solution standard. Appuyer
sur YES.

Régler la valeur à l’aide des touches >, ^ et ENT

Sélectionner le digit clignotant avec > et augmenter
sa valeur à l’aide de ^.

Valider avec ENT

WAIT s’affiche brievement CAL.END apparaît

L’étalonnage est terminé. Replonger le capteur
dans le proceé. Appuyer sur YES pour repasser en
mode mesure.

Note: voir section 4-5.