Annexe 1 Tableau de réglage utilisateur PH/ORP

App.1-6

IM 12A01A02-12F-E

Paramètre

Valeurs par

défaut

Valeurs applicables Réglages *Réf.

Display

setup

Dual

display

1st row

PH1

PH1, PH2, Calculated.

Redundant

5.7.1

2nd row

PH2

PH1, PH2, Calculated.

Redundant, Empty

Individual

display

Main

display

PH1 (2)

1st line

pH1 (2)

pH1 (2),

Temperature1 (2),

ORP1 (2), rH1 (2),

Concentration1 (2)

2nd line

Temperature1

(2)

pH1 (2), Tempera-

ture1 (2), ORP1 (2),

rH1 (2), Concentra-

tion1 (2), Empty

3rd line

Empty,

(ORP1)

Additional text

(blank)

(Max. 12 characters)

Trend

Trend

Graph

Screen:

1st Trend

pH1

pH1 (2),

Temperature1 (2),

ORP1 (2), rH1 (2),

Concentration1 (2),

Calculated, Redun-

dant, Empty

5.7.2

2nd Trend

Temperature1

3rd Trend

Empty,

(ORP1)

X-axis:

Timing

TIme Span (X)=

1 hour

15 min., 30 min., 1

hour, 2 hours, 4 hours,

8 hours, 24 hours, 7

days, 14 days

Y-axis:

Limits

pH

low

0.00 pH

-2 à 20 pH

high

14.00 pH

Tempera-

ture

low

0.0 ºC

-30 à 140 ºC

high

100.0 ºC

ORP

low

-500 mV

-1500 à 1500 mV

high

500 mV

rH

low

0 rH

0 à 100 rH

high

50 rH

Calculate

(Differential)

low

-1pH, -500mV -20 à 20 pH,

-1500 à 1500 mV

high

1 pH, 500 mV

Calculate

(Average)

low

0 pH, -500mV -20 à 20 pH,

-1500 à 1500 mV

high

14pH, 500mV

Concentra-

tion

low

0 %, mg/L,

g/L, ppm

0 а 100 %,

0.000 а 9999 mg/L,

g/L, ppm

high

100.0 %,

mg/L, g/L,

ppm

Auto

Return

Auto Return

10min

Dissable, 10min,

60min

5.7.3

Adjust

contrast

Contrast:

0

-5 à +5

5.7.4

Calculated

data setup

Function

Differential

Differential, Average

5.8