App.1-5 – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 Manuel d'utilisation

Page 44

Advertising
background image

Annexe 1 Tableau de réglage utilisateur PH/ORP>

App.1-5

IM 12A01A02-12F-E

Paramètre

Valeurs par

défaut

Valeurs applicables Réglages *Réf.

Advanced

setup

Settings

No action

No action, Load

factory settings, Save

user settings, Load

user settings

5.6.1

Tag

FLXA21-PH

(Max. 12 characters)

5.6.2

Passwords Execute:

(blank)

(Max. 8 characters)

5.6.3

Commissioning:

(blank)

(Max. 8 characters)

Date/Time

Format

YYYY/MM/

DD

YYYY/MM/DD,

MM/DD/YYYY,

DD/MM/YYYY

5.6.4

Date

Time

Communication

HART

None, HART, PH201G

5.6.5

HART

Network address

0

0 à 15

PV

(pH1)

SV

Temperature1 pH1 (2), Tempera-

ture1 (2), Ref.imped-

ance1 (2), ORP1 (2),

rH1 (2), Calculated,

Redundant, Empty

TV

Empty, pH2

FV

Empty,

Temperature2

PH201G

Hold contact:

Enabled

Disabled, Enabled

Fail contact:

Fail+Warn

Fail+Warn, Fail only,

Disabled

Wash contact

Enabled

Disabled, Enabled

Interval time

6.0 h.

0.1 à 36.0 h.

Wash (Meas-

ure) time

0.5 min.

0.1 à 10.0 min.

Recovery time

0.5 min.

0.1 à 10.0 min.

Manual Wash

Enabled

Disabled, Enabled

Imp2 Wash

Disabled

Disabled, Enabled

Continuous

Wash

Disabled

Disabled, Enabled

Factory setup

5.6.6

PH

Advertising