En-9 paj-n series, 3 communication – Ohaus PAJ GOLD PLUS PRECISION JEWELRY BALANCES Manual multi Manuel d'utilisation

Page 10

Advertising
background image

EN-9 PAJ-N

Series

InCal Models

MODEL PAJ603CN

PAJ1003CN

Max

600ct

1000ct

Min 0.2ct
d= 0.001ct
e= 0.01ct
Approval Class

II

Tare Range

To capacity by subtraction

Stabilization 3

Seconds

Span cal mass 50g or 100g 100g or 200g
Lin. cal masses 50g, 100g

200g, 500g

Pan dia.

∅3.5in / ∅9cm

Net Weight

10.1 lb / 4.6kg


7.3 Communication

7.3.1 Commands
The RS232 Interface allows a computer to control the balance as well as receiving data such as the displayed
weight. The balance will return “ES” for invalid commands.

Command Function

IP

Immediate Print of displayed weight.

P

Print displayed weight (uses stable on/off menu setting).

CP Continuous

Print.

xP

Interval Print x = Print Interval (1-3600 sec)

T

Same as pressing Zero Key.

ON Turns

balance

ON.

OFF Turns

balance

OFF.

PSN Show

Serial

Number.

PV

Version: Print product name, software revision and LFT ON (if LFT is set ON).

PU

Print current mode/unit

x#

Set PC ref wt (x) in grams

P#

Print PC ref wt

x%

Set % ref wt (x) in grams

P%

Print % ref wt


7.3.2 RS232 (DB9) Pin Connections
Pin 2: Balance transmit line (TxD)
Pin 3: Balance receive line (RxD)
Pin 5: Ground signal (GND)
Pin 7: Clear to send (hardware handshake) (CTS)
Pin 8: Request to send (hardware handshake) (RTS)

CTS

RTS

5

9

1

6

RxD

TxD

GND

Advertising