Specifications – Ohaus B250P_S_AS SCALE BASES Manuel d'utilisation

Page 9

Advertising
background image

9

NIST Handbook 44 Class III
14° to 104°F/-10° to 40° C

Model

NTEP No. 95-151

B10S

n

max

5000

B25S

n

max

5000

B50S

n

max

5000

B100S

n

max

5000

B150S

n

max

3000

B250S

n

max

5000

*

In type approved applications, readability is limited to capacity/n

max.

Specifications are given as a percent of the rated load.

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

MODEL

MODEL

MODEL

MODEL

MODEL

B10S

B10S

B10S

B10S

B10S

B25S B50S B100S B150S B250S

B25S B50S B100S B150S B250S

B25S B50S B100S B150S B250S

B25S B50S B100S B150S B250S

B25S B50S B100S B150S B250S

Capacity

(lb)

25

50

100

200

300

500

(kg)

10

25

50

100

150

250

Readability

(lb)

0.001

0.002

0.005

0.01

0.01

0.02

(Non type approved)

(kg)

0.0005

0.001

0.002

0.005

0.005 0.01

Readability

*

(lb)

0.005

0.01

0.02

0.05

0.1

0.1

(Type aproved)

(kg)

0.002

0.005

0.01

0.02

0.05 0.05

Repeatability (Std. dev.)

+ 0.01%

Linearity

+ 0.02%

Safe overload capacity

150%

Load cell excitation (V dc)

5 to 15

Operating temperature range

14° to 104°F/-10° to 40° C

Scale base size

12 x12 x 3.4/

18 x 18 x 3.6/

24 x 24 x 5.8/

(WxDxH) (in/cm)

30 x 30 x 8.6

45 x 45 x 9.1

60 x 60 x 14.8

Scale base construction
Platform

Stainless steel

Frame

Painted aluminum

Painted steel

Net weight (lb/kg)

10/4.5

35/16

71/32

Shipping weight (lb/kg)

13/6

42/19

82/37

Advertising