3.2 Remote control commands

768 KF Oven

27

.Rate:

Rate in mm/s

.Pos:

Position moved to with &Assembly.Boat $G, in mm. The inputted position is
moved to irrespective of the limits set in &Assembly.Boat.SetPos.

.SetPos:

Limits for the movement of the boat in automatic determinations and for
manual operation.

.InPos:

Inner stop in mm.

.OutPos:

Outer stop in mm.



3.2.2.29

Assembly.Pump

$G, $S

Switch pump on/off.


3.2.2.30

Assembly.Outputs.SetLines

$G

Assembly.Outputs.SetLines.L1

active, inactive, pulse, OFF

up to .L8

Assembly.Outputs.ResetLines

$G

Setting of the output lines of the "Remote" socket.
.SetLines:

All lines are set with $G.

.SetLines.L1: Setting of line 1. "active" sets a static signal (0 V), "inactive" resets the signal

(+5 V) , "pulse" means a pulse of length of ca. 150 ms and with "OFF" the
corresponding line is not controlled.
Assignment of the lines, see 3.2.2.20.

up to .L8

.ResetLines: Lines L1...L8 are set to the inactive condition (+5 V).


3.2.2.31

Setup.IdReport

ON, OFF

Switch output of the report identification on/off.


3.2.2.32

Setup.Keycode

ON, OFF

With ON the key code of a key pressed on the KF Oven is outputted. The key code
comprises 2 ASCII characters see below. Space (ASCII 32) and # are sent as introducers.
Example : #11



3.2.2.33

Setup.Tree.Short

ON, OFF

Setup.Tree.ChangedOnly

ON, OFF

Definition of the response to an inquiry with $Q.
.Short:

With ON all path names are sent, but only with the requisite number of
characters. Requisite number of characters means the path is unequivocally
characterised (characters printed in boldface in the instructions for use).