3.1 General rules

768 KF Oven

19


RS send errors:

E40

DSR=OFF. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?

E41

DCD=ON. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?

E42

CTS=OFF. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?

E43

The transmission has been interrupted with XOFF for at least 3
s.
Exit: Send XON or <QUIT>.

E44

The RS parameters are no longer the same for both devices.
Reset.

E45

The receive buffer contains an incomplete command (LF

missing). Sending from the instrument is therefore blocked.
Exit: Send LF or <QUIT>.



Device-specific errors:

E26

Manual stop.
Exit: The error message disappears on the next start.

E135

Check temperature sensor of sample temperature.
Exit: Rectify fault.

E154

Sample temperature not OK.
Exit: The error message disappears when the fault is rectified
or &M $S.

E163

Gas flow too low.
Exit: The error message disappears when the fault is rectified
or &M $S.

E164

The attached titrator is not conditioned.

Exit: The error message disappears when the fault is rectified
or &M $S.

E165

The oven temperature is higher than 360 °C.
Exit: The error message disappears when the fault is rectified.

E168

The temperature sensor for the oven temperature is not OK.
Exit: Rectify fault.

E169

The gas flow sensor is not OK (flow >500 mL/min).
Exit: Rectify fault. Continuation is possible only when the
minimum gas flow is set to 0 mL/min.