3.2 Remote control commands

768 KF Oven

23

3.2.2 Description of the remote control commands


3.2.2.1

Mode

$G, $S

Start and stop ($G, $S) of the automatic determination.


3.2.2.2

Mode.Temp

50...300

Entry of the sample temperature in °C.


3.2.2.3

Mode.Gas.UnitFlow

mL/min, L/h

Mode.Gas.MinFlow

0...5...999

.UnitFlow:

Selection of the unit for display of the gas flow.

.MinFlow:

Setting of the minimum gas flow in the unit selected above. If the flow is less

than the minimum gas flow, error E163 appears (gas flow too low).



3.2.2.4

Mode.Gas.Type.Select

air, N2, other

Mode.Gas.Type.OtherFac

0.001...1...9999

.Select:

Selection of the gas.

.Factor:

If "other" has been selected above, setting of the factor for measurement of

the gas flow, factors see page 11.



3.2.2.5

Mode.Gas.PurgeTime

0...99999

Mode.Gas.CondTime

0...99999

.PurgeTime: Purge time with valve to "Purge". Entry in s.
.CondTime:

Conditioning time with valve to "Transfer". Entry in s.



3.2.2.6

Config.OvenSet.AutoPrep

ON, OFF

ON means: Automatic heating and preparation of the oven after switching on.


3.2.2.7

Config.OvenSet.ValveControl

ON, OFF

Config.OvenSet.StartCond

ON, OFF

Config.OvenSet.TempLimit

1...5...100

Config.OvenSet.TempCorr

0...

±99.9

• ValveControl:

ON means automatic switching of the valve from PURGE to TRANSFER at the end of the
determination.

• StartCond:

ON means: Insert sample boat during automatic determination only if the connected
device is conditioned (input line pin 12 of the socket "Remote" active).

• TempLimit:

Starting a determination is possible only if the current temperature is in the range of

±X

°C from the sample temperature.

• TempCorr:

Correction of the sample temperature according to individual sample and Oven
requirements.