Metrohm 795 KFT Titrino Manuel d'utilisation

Page 85

Advertising
background image

3.2 Remote control commands

795 KFT Titrino

81

3.2.2.68. Info.ActualInfo.Assembly.Counter.V read

only

Info.ActualInfo.Assembly.Counter.Clear $G

$Q sends the volume. With the function &Info.Assembly.Counter.Clear $G, the

volume counter is set to zero.

3.2.2.69. Info.ActualInfo.Assembly.Meas read

only

$Q sends the current measured value from the assembly.

3.2.2.70. Info.ActualInfo.Titrator.CyclNo read

only

Info.ActualInfo.Titrator.V read

only

Info.ActualInfo.Titrator.Meas read

only

Info.ActualInfo.Titrator.dVdt read

only

Info.ActualInfo.Titrator.dMeasdt read

only

Info.ActualInfo.Titrator.dMeasdV read

only

Info.ActualInfo.Titrator.T read

only

$Q sends the current values in the following formats:

KFT
CyclNo 127
V(mL) 1.2345
Meas:

Ipol (mV)
Upol (µA)

–345.6

–12.5

dVdt (µL/s)

2.5142

dMeasdt

Ipol (mV/s)
Upol (µA/s)

0.7957

0.7957

dMeasdV (mV/µL)

10.6326

NV: Not Valid.

OV will be sent for "overrange".

A time frame can be set up from the cycle number and the cycle time (see

3.2.2.75). The cycle number is set to 0 at the start of a method and it is in-

cremented until the end of the method.

3.2.2.71. Info.ActualInfo.MeasPt.Index read

only

Info.ActualInfo.MeasPt.X read

only

Info.ActualInfo.MeasPt.Y read

only

Info.ActualInfo.MeasPt.Z1 read

only

Info.ActualInfo.MeasPt.Z2 read

only

Info.ActualInfo.EP.Index read

only

Info.ActualInfo.EP.X read

only

Info.ActualInfo.EP.Y read

only

$Q sends the last entry into the measuring point list (.MeasPt) or the last entry

into the list of EP's.