Chapter 7: Modifying Your Protector Premier
Product Service 1-800-522-7658
33
Original instructions
Installing Guardian™ Digital Airflow
Monitor or Guardian™ Airflow Monitor
The Guardian Airflow Monitors continuously monitor face velocity through the
fume hood opening. The fume hood right corner post is factory prepared to
mount either monitor. Contact Labconco Customer Service to order.
Distillation Grids – Field Installation
The distillation grids have been strategically placed with the vertical rod center
lines in front of the lower baffle and middle baffle. The distillation grids allow
the hood user to mount glassware, motors, stirrers, and other apparatus (not
available on Pass-Through Fume Hood). Contact Labconco Customer Service for
ordering information.
Figure 7-6
Sash Stop Kit – Field Installation
(P/N 9410300)
The sash stop kit restricts how far a vertical-rising sash may be opened. This
small plastic device may be easily field installed on the fixture corner post of any
fume hood.
Upper Bypass for Premier Field Installation
The upper bypass prepares the Premier for use with variable air volume (VAV)
systems and allows the hood to use less (CFM) as detailed in Chapter 2. The part
numbers are listed below:
4' Premier Upper Bypass, P/N 9409214
5' Premier Upper Bypass, P/N 9409215
6' Premier Upper Bypass, P/N 9409216
8' Premier Upper Bypass, P/N 9409218