Chapter 6: Troubleshooting

32

Original Instructions

PROBLEM CAUSES CORRECTIVE

ACTION

Reduced flow at
dispense valve/gun
and storage tank
slow to fill

Restricted tube in the feed line or
polishing system

Inspect polishing loop tubing
for any restrictions

Insufficient feed water pressure and
flow rate

Pressure must be a minimum
of 30 psi at a flow rate of at
least 6 liters per minute
Check inline strainer for
obstruction

Partial open feed valve

Open feed valve fully

Plugged prefilter and/or carbon filter Replace

filters

Water temperature less than 25°C

Install water temperature
mixing valve

Kinked or restricted tubing in system Trace restriction and remove

or straighten tube

Decrease in RO back pressure

Increase back pressure relief
valve to 150 psi

Water quality
display acting
erratically

New cartridges installed in system

Purge all of the air out of the
polishing system

Air trapped in conductivity cell

Purge all of the air out of the
RO system

Polishing loop conductivity cell not
connected to wiring harness

Reconnect cell to wiring
harness

Conductivity cell is defective

Replace cell

Display reads 000

Conductivity sensor disconnected
from harness

Connect sensor to harness
from circuit board

Air in conductivity sensor

Purge unit until air is expelled
by opening the dispensing
valve