Chapter 2: Prerequisites

7

Original Instructions

There are two drain lines. One is from the tank overflow drain and the other is from the RO for
reject water. The drain must be located lower and must be able to handle a minimum of 2-gal/min
water flow.

IMPORTANT NOTE: System performance, membrane, and cartridge life span are directly
related to the feed water. It is important to establish feed water pressure, and flow rate, before
installing and operating the unit. If your feed water, or feed water quality has not been tested,
contact Labconco at 1-800-821-5525 or 1-816-333-8811.

Space Requirements

Wall mounted units require a minimum of 6" clearance on the bottom for cartridge, and membrane
installation/removal.

Refer to APPENDIX B: WATERPRO RO SYSTEM DIMENSIONS for dimensional drawings of
the WaterPro RO System.