Dynalab 3999-0000 Manuel d'utilisation

Page 7

Advertising
background image

[ 7 ]

LIST OF SPARE PARTS

The following components are available from most laboratory suppliers. In case of dif ficulty please contact Bibby Sterilin Ltd.

Component

Catalogue No.

Component

Catalogue No.

Support stand

D4000/S

P.T.F.E. washer

QW18/7

Boiler

D4000/B

Reservoir level control

WLS

Condenser WC48/M2

Boiler thermostat

WT4

Heating element 240V

D240H

Condenser thermostat

W14TS

Heating element 220V

D220H

Reservoir level control pressure switch

DPS

Hose kit

W4000/HK

On/Off switch

D4000/SW

Gasket kit

D4000/GK

Boiler O rings

Top

WB48/01

Boiler thermostat tube

WTT48

Bottom

QR38/24

Condenser thermostat tube

A515414

Heater sealing ring

QR38/24

Plastic connecting caps

QC18/11

Silicone rubber rings

QR18/7 and
QR18/10

FAULT FINDING
In the event of operating difficulties with your Distinction Water Still, please consult the following notes. If these fail to
identify and remedy the fault you are advised to contact your supplier or the Technical Service Department of Bibby Sterilin
Ltd.

Symptom

Cause

Remedy

Heater not working

a) Burnt out heater

Replace heater

b) Failed electricity supply

Restore supply

c) Thermostat operated

Identify and rectify the
cause and reset thermostat

Water level in boiler

a) Supply of cooling water is

Adjust the supply to

is too low - heater exposed

insufficient

approximately 60 l/hr

b) Boiler drain stopcock left open

Close stopcock

Water level in boiler

a) Flow of cooling water is too high

Adjust the supply to

is too high - e.g. water surging

approximately 60 l/hr

into condenser

b) Drain is restricted

Ensure drain pipe falls freely
from Still with no kinks

Distillate temperature is

Supply of cooling water is

Adjust the supply to

too high - e.g. >50°C

insufficient

approx 60 l/hr

Water in boiler is “pumped”

a) Tubing from condenser

Ensure tubing falls freely

out of the boiler to drain

distillate outlet to the reservoir

from Still with no kinks

is constricted

b) Mounting for condenser

Remove obstruction

thermostat is blocked

c) Supply of cooling water

Adjust the supply to

is insufficient

approximately 60 l/hr

Distillate quality is poor*

Boiler heavily scaled

Clean boiler

Output less than 4 l/hr

Mains electricity supply below

nominal voltage

*Distillate quality when measured by pH or conductivity is gr eatly affected by temperature and the presence of absorbed
carbon dioxide.