FAULT FINDING

In the event of operating difficulties with your Merit Still, please consult the following notes. If these fail to identify and

remedy the fault, then you are advised to seek the help of your supplier or the Technical Services Department of

Bibby Sterilin Ltd.

Note – Fault finding should only be carried out by suitably qualified people.

Symptom

Cause

Remedy

1. Water level in boiler is too

a) Supply of feed/cooling water

a) Increase flowrate of water to

LOW. e.g. – heater exposed.

is insufficient.

approx. 60 litres/hour.

b) Stopcock on boiler inadvertently

b) Close stopcock.

left open.

2. Water level in boiler is too

a) Supply of feed/cooling water

a) Reduce flowrate of water to

HIGH. e.g. – boiling water

is excessive.

approx. 60 litres/hour.

surging into condenser.

b) Flow of drainage water

b) Ensure drainage tubing falls freely.

is constricted.

3. Water in boiler is “pumped”

a) Tubing from condenser distillate

a) Ensure tubing falls freely to

out of boiler to drain.

outlet to reservoir is constricted.

reservoir without kinks or bends.

b) Vent on condenser distillate

b) Remove obstruction.

outlet is blocked.

c) Supply of feed/cooling water

c) Increase flowrate of water to

is insufficient.

approx. 60 litres/hour.

4. Distillate temperature is

Flow of cooling water is insufficient.

Increase flowrate of water to

high, e.g. – above 50°C.

approx. 60 litres/hour.

5. Distillate rate less than

Mains voltage low.

4 litres/hour.

Excessive cooling water flow.

Reduce to 60 litres/hour.

6. Distillate quality poor*.

Boiler heavily scaled.

Clean boiler.

7. Heater not working.

a) Burnt-out heater.

a) Replace heater.

b) Mains electricity fuse blown.

b) Replace fuse.

c) Thermostat operated.

c) Reset.

* Distillate quality when determined by pH or electrical conductivity is greatly affected by temperature and the

presence of absorbed carbon dioxide.