Important Control Solution Test Information

• Use only OneTouch

®

Ultra

®

Control Solution.

• Check the expiration date on the control solution vial. Record

the discard date (date opened plus three months) on the vial
label. Do not use after expiration or discard date, whichever
comes first.

• Control solution, meter, and test strips should be at room

temperature (20–25°C/ 68–77°F) before testing.

• Shake the vial, discard the first drop of control solution, and wipe

off the tip to ensure a good sample and an accurate result.

• Store control solution tightly closed at temperatures below 30°C

(86°F). Do not refrigerate.