Caractéristiques techniques, Copper specifications – EXFO MaxTester 635 Manuel d'utilisation

Page 339

Advertising
background image

Caractéristiques techniques

Testeur de cuivre et de services multiples VDSL2/ADSL2+

331

COPPER SPECIFICATIONS

a, b, c

Transmitter characteristics

Frequency range (200 Hz to 20 kHz)

Frequency resolution

1-Hz steps

Frequency uncertainty (accuracy)

±(50 ppm + 1 Hz)

Level range (dBm)

–20 to 10 at 600

1

Level resolution

0.1 dB

Level uncertainty (accuracy)

±1 dB

Impedance (

1)

600

Frequency range (20 kHz to 2.2 MHz)

Frequency resolution

1-kHz steps

Frequency uncertainty (accuracy)

±(50 ppm + 100 Hz)

Level range (dBm)

–20 to 10 at 100

1

Level resolution

0.1 dB

Level uncertainty (accuracy)

±1 dB

Impedance (

1)

100, 120, 135, 150

Frequency range (2.2 MHz to 30 MHz)

Frequency resolution

1-kHz steps

Frequency uncertainty (accuracy)

±(50 ppm + 100 Hz)

Level range (dBm)

–20 to 0 at 100

1

Level resolution

0.1 dB

Level uncertainty (accuracy)

±1 dB

Impedance (

1)

100, 120, 135, 150

Receiver characteristics

Reception frequency range

200 Hz to 20 kHz
20 kHz to 30 MHz

Frequency uncertainty range (accuracy)

±(50 ppm + 1 digit)

VF reception level range (dBm)

–90 to 15 at 600

1

VF level uncertainty (accuracy)

200 Hz to 20 kHz
–90 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB

WB reception level range (dBm)

–90 to 15 at 100

1 and 135 1

WB level uncertainty (accuracy)

20 kHz to 2.2 MHz
–90 dBm to –50 dBm,

uncertainty (accuracy)

±2 dB

–50 dBm to 15 dBm,

uncertainty (accuracy)

±1 dB

2.2 MHz to 30 MHz
–90 dBm to –50 dBm,

uncertainty (accuracy)

±2 dB

–50 dBm to 15 dBm,

uncertainty (accuracy)

±1 dB

Impedance (

1)

100, 120, 135, 150, 600

POTS dialer

DTMF

0 – 9, #, *

Phonebook

25 entries

Digital multimeter (DMM)

Test type

Snapshot and continuous

Impedance selection (for voltage measurement)

100 k

1, 1 M1

Measurement

Range

Resolution

Uncertainty (accuracy)

DC voltage

0 to 400 V

0.1 V for 0 to 99.9 V
1 V for 100 V to 400 V

±(|1 %| + 0.5 VDC)

AC voltage

0 to 280 Vrms

0.1 V for 0 to 99.9 V
1 V for 100 V to 280 V

±(1 % + 0.5 VAC)

Isolation resistance
(stress/leakage)

0 to 1 G

1, auto-ranging

1 k

1 to 99 M1

100 M

1 to 999 M1

Three digits

±(2 % + 1 digit)
±(5 % + 1 digit)

Resistance

0 to 100 M

1

0 to 999

1

1 k

1 to 100 M1

Three digits

±(1 % + 5

1)

±(2 % + 1 digit)

Capacitance

0.1 nF to 2 μF

Four digits

±(2 % + 50 pF)

DC current

0 to 110 mA

0.1 mA

±(|2 %| + 1 mA)

AC current

0 to 110 mA

0.1 mA

±(

|2 %|

+ 1 mA)

d

Station Ground

0 to 1 M

1

0 to 999

1

1 k

1 to 1 M1

Up to three digits

±(1 % + 3

1)

±(2 % + 1 digit)

NOTES

a. Subject to change without notice.
b. Typical, at 23 °C ± 3 °C, on batteries, with no type B USB connection.
c. Specifications based on 24 AWG (PE 0.5 mm) cabling.
d. From 10 mA to 110 mA.

Advertising