Inserting the seals, Iii exchanging the filling tube, Iv exchanging the tip cone – Eppendorf Reference Manuel d'utilisation

Page 30: Exchanging the filling tube, Exchanging the tip cone, Eppendorf reference® – part b – maintenance

Advertising
background image

101

Inserting the seals

Pull new seal off the pin and push screw, spring, seal (with plastic part first) in that
order onto side B of the wrench as shown in the figure and screw lightly into tip
cone. Do not tighten too much. Assemble pipette again.

10 – 100 μL, 50 – 200 μL and 50 – 250 μL (Fig. 3, 4, 5)
Filling tube (12)
– Unscrew lower part of the pipette.
– Push filling tube (12) (and damping tube (14) for 50 – 250 μL) out of the tip

cone from below with the wire punch. Push in new tube from above.

After the ejector sleeve has been pulled off, the tip cone together with the
seal (11) can be loosened with the wrench (A, narrow opening) and exchanged
(together with the seal. See Fig. 2 – 6).

Tighten the tip cone:
Place wrench with the wide opening (A, lettering facing the pipette tip) onto the tip
cone and tighten until it locks into position by turning half a revolution.
The wrench is designed in such a way that the tip cone cannot be tightened too
much.

Caution
After exchanging parts or completing other maintenance, always check that the
pipette functions correctly (see Part A, Section 5 and 7 of this Manual).
If a problem cannot be solved with the aid of the recommendations above, please
return the eppendorf Reference to your eppendorf appointed dealer.

$

%

'

&

III

Exchanging the filling tube

IV

Exchanging the tip cone

Eppendorf Reference® – Part B – Maintenance