Chapter 6: Troubleshooting
Product Service 1-800-522-7658
26
26
PROBLEM
CAUSE
CORRECTIVE ACTION
Washer will not
drain
Clogged drain
Drain hose too high
Or hanging too low
Clean out drain.
Reroute drain hose
Filter blocked
Check filter in washer and clean as needed.
Drain valve or air
valve not operating
Check operation.
Washer overfills
Defective fill switch
Check operation.
Selection buttons
on control panel
inoperable or
controls
malfunction
Microprocessor
memory error
Perform "Factory Reset." Press and hold MENU
button and simultaneously turn on power switch.
Release MENU button and display will show the
diagnostics screen. Press MENU until Factory
Reset is displayed. Press SELECT. Press MENU
Return to Menu is displayed. Press SELECT.
Key pad defective
Replace key pad.
Water Low Alarm
Improper water
supply
Make sure water is turned on.
Water pressure must be at least 8 psi
(55kPa) and flow must be at least 1.25 gallons
(4.7 liters)/minute.
At least .67 gallons (2.5 liters) must be available
if using pure water.
Defective fill valve
Clean screens inside the inlet fittings of the water
valve.
Check voltage to valve.
Replace valve if it does not operate with proper
voltage.
Restricted flow
Check for kinks in water hoses in base of washer
to water inlet fitting on side of tank.
Pipettes do not dry
Dry time to short
Increase dry time.
Defective air valve
Check operation in diagnostics
Check voltage to valve
Temp Fail Alarm
Temperature sensor
failure
Replace valve if it does not operate with proper
voltage.
Check connection.
Replace sensor