Caractéristiques techniques

37

General Specifications

Size (H x W x D)

18.5 cm x 10.0 cm x 5.5 cm

(7

1

/

4

in x 4 in x 2

1

/

8

in)

Weight

0.4 kg

(0.9 lb)

Temperature

operating

-10 °C to 50 °C

(14 °F to 122 °F)

storage

-40 °C to 70 °C

(-40 °F to 158 °F)

Relative humidity

0 % to 95 % non-condensing

Standard Accessories

User guide, Certificate of Calibration, instrument stickers in six languages, AC
adapter, connector adapter (FOA-XX), three AA batteries, wrist strap, alcohol
cleaning pads.

Safety

21 CFR 1040.10 and IEC 60825-1:1993+A1:1997+A2:2001:
CLASS 1M LASER PRODUCT

Model

FOT-302

FOT-302X

Power meter port

2

Ge

GeX

Power range (dBm)

3

10 to -60

26 to -50

Range displayed (dBm)

Down to -65

Down to -50

Number of calibrated wavelengths

4

10

10

Power uncertainty

5

± 5 % ± 1 nW

± 5 % ± 10 nW

Resolution (dB)

0.01

6

0.01

7

Automatic offset nulling

8

Yes

Yes

Warmup time (s)

9

0

0

Display units

dB/dBm/W

dB/dBm/W

Automatic wavelength recognition

10

Yes

Yes

Screen refresh rate (Hz)

3

3

Tone detection (Hz)

270, 1 k, 2 k

270, 1 k, 2 k

Battery life (hours) (typical)

260

260

Warranty and recommended

3

3

calibration interval (years)

Notes
1. Guaranteed unless otherwise specified.
2. All specifications valid at 1550 nm and 23 °C ± 1 °C, with an FC connector.
3. In CW mode; sensitivity defined as 6 x rms noise level.
4. Wavelengths: 830 nm, 850 nm, 980 nm, 1300 nm, 1310 nm, 1450 nm, 1490 nm,

1550 nm, 1590 nm and 1625 nm.

5. Traceable to NIST; FOT-302X: up to 20 dBm.
6. From 10 dBm to -50 dBm.
7. From 26 dBm to -35 dBm.
8. Power of > -40 dBm for FOT-302, and of > -25 dBm for FOT-302X.

9. For ± 0.05 dB, for temperatures > 18 °C.
10. At 850 nm, 1300 nm, 1310 nm, 1490 nm, 1550 nm and 1625 nm;

power > -50 dBm for FOT-302, and > -40 dBm (typical) for FOT-302X.

11. All specifications valid at 23 °C ± 1 °C, with an FC connector.
12. rms for lasers and -3 dB width for LEDs; typical values for LEDs.
13. After 15 minutes warmup; expressed as ± half the difference between the maximum

and minimum values measured during the period, with an APC connector on the power
meter.

14. Typical autonomy in Auto mode.

Model

11

23BL

234BL

235BL

12D

01-VCL

Central wavelength (nm)

1310 ± 20

1310 ± 20

1310 ± 20

850 ± 25

850 ± 20

1550 ± 20

1550 ± 20

1490 ± 10

1300 +50/-10

1625 ± 15

1550 ± 20

Spectral width (nm)

12

≤ 5

≤ 5

≤ 5

50/135

≤ 1

Output power (dBm)

≥ 1/≥ 1

≥ 1/≥ -3/≥ -5

≥ 1/≥ -4.5/≥ -3

≥ -18/≥ -18 (62.5/125 μm)

≥ -3 (50/125 μm)

Power stability (dB)

13

8 hours

± 0.10

± 0.10

± 0.10

± 0.10

± 0.25

Battery life (hours)

14

120

100

120

120

250

Automatic wavelength recognition

Yes

Yes

Yes

Yes

Yes

Tone generation (Hz)

270, 1 k, 2 k

270, 1 k, 2 k

270, 1 k, 2 k

270, 1 k, 2 k

270, 1 k, 2 k

Warranty and recommended calibration interval (years)

3

3

3

3

3