Metrohm 768 KF Oven Manuel d'utilisation

Page 31

Advertising
background image

3.2 Remote control commands3.2 Remote control commands

768 KF Oven

28

.ChangedOnly: Sends only those paths and their values which have once been edited. All

path names are sent in absolute terms, i.e. from the root.



3.2.2.34

Setup.Trace

ON, OFF

The KF Oven reports when a value has been confirmed with <ENTER> at the KF Oven.
Space (ASCII 32) and & are sent as introducers.
Example: &Mode.Temp"100"


3.2.2.35

Setup.Lock.Keyboard

ON, OFF

Setup.Lock.Config

ON, OFF

Setup.Lock.Parameter

ON, OFF

Setup.Lock.Heater

ON, OFF

Setup.Lock.Pump

ON, OFF

Setup.Lock.Valve

ON, OFF

Setup.Lock.Boat

ON, OFF

Setup.Lock.Display

ON, OFF

ON means locking of the corresponding function:
.Keyboard:

Locking all keys at the KF Oven.

.Config:

Locking the <CONFIG> key.

.Parameter: Locking the <PARAM> key.
.Heater:

Locking the <HEATER> key.

.Pump:

Locking the <PUMP> key.

.Valve:

Locking the <VALVE> key.

.Boat:

Locking the <BOAT IN> and <BOAT OUT> keys.

.Display:

Locking the display, it is not operated by the device program of the KF Oven
and can be written to by the computer.



3.2.2.36

Setup.TController.InitHeatFactor

0...100...200

Setup.TController.AddHeatFactor

0...100...200

Factors for the temperature controller, see page 12.


3.2.2.37

Setup.SendMeas.SendStatus

ON, OFF

Setup.SendMeas.Interval

1...4...16200

Setup.SendMeas.Meas.CyclNoON, OFF

Setup.SendMeas.Meas.SampleTemp

ON, OFF

Setup.SendMeas.Meas.OvenTemp

ON, OFF

Setup.SendMeas.Meas.GasFlowON, OFF

.SendStatus: ON means automatic transmission of measured values (see below) at the

inputted time interval.

.Interval:

Time interval for the transmission of associated measured values defined
below.


Selection of the values which are transmitted at the set time interval.
.CyclNo:

Cycle number. Together with the cycle time (3.2.2.24), a time frame can be
generated. The other data are part of the corresponding cycle number. The
cycle number is always zeroed at the start and termination of the automatic
determination.

.SampleTemp: Associated sample temperature.
.OvenTemp: Associated oven temperature.
.GasFlow:

Associated gas flow.


Format of the transmitted values (examples):
.CyclNo

Cycle number

127

.SampleTemp Sample temperature in °C

150.0

.OvenTemp Oven temperature in °C

170.0

.GasFlow

Gas flow in mL/min or L/h

100.5

Advertising