Sira certification service, Inficon - manuel de l’utilisateur extrima – INFICON Extrima Ex-certified Hydrogen Leak Detector Manuel d'utilisation

Page 51

Advertising
background image

51

FR

INFICON - Manuel de l’utilisateur Extrima

Sira Certification Service

Rake Lane, Eccleston, Chester, CH4 9JN, England
Tel:

+44 (0) 1244 670900

Fax:

+44 (0) 1244 681330

Email:

[email protected]

Web:

www.siracertification.com

C E R T I F I C A T I O N

SCHEDULE

EC TYPE-EXAMINATION CERTIFICATE

Sira 07ATEX2117X

Issue 3

This certificate and its schedules may only be

reproduced in its entirety and without change.

Page 3 of 3

Form 9400 Issue1

Variation 2 - This variation introduced the following changes:
i.

The LCD module for the Extrima® Hydrogen Leak Detector was modified and now includes

components with a surface area of less than 20 mm².

i.

The bill of material drawings, KK1012-BOM-1H-CERT and KK1018-BOM-R7-CERT, were amended to:

• Bring them into line with Sira report number R20666A/01.

• Remove the manufacturer’s name from the specification of various safety resistors.

14

DESCRIPTIVE DOCUMENTS

14.1

Drawings
Refer to Certificate Annexe.

14.2

Associated Sira Reports and Certificate History
Issue

Date

Report no.

Comment

0

10 October 2007

R52A16411B

The release of the prime certificate.

1

18 December 2009

R20666A/00

The introduction of Variation 1

(Note: the date was revised

by Issue 3 to correct a typographical error)

.

2

30 April 2010

R20666A/01

Issued to allow report R20666A/01 to replace report

R20666A/00

3

20 October 2010

R23373A/00

R23526A/00

The introduction of Variation 2.

15

SPECIAL CONDITIONS FOR SAFE USE (denoted by X after the certificate number)

15.1

As aluminium is used at the accessible surface of this equipment, in the event of rare incidents, ignition

sources due to impact and friction sparks could occur. This shall be considered when the Extrima®

Hydrogen Leak Detector is being used in locations that specifically require group II, category 1

equipment.

16

ESSENTIAL HEALTH AND SAFETY REQUIREMENTS OF ANNEX II (EHSRs)
The relevant EHSRs that are not addressed by the standards listed in this certificate have been

identified and individually assessed in the reports listed in Section 14.2.

17

CONDITIONS OF CERTIFICATION

17.1

The use of this certificate is subject to the Regulations Applicable to Holders of Sira Certificates.

17.2

Holders of EC type-examination certificates are required to comply with the production control

requirements defined in Article 8 of directive 94/9/EC.

17.3

The battery pack shall be constructed from three, series connected SAFT type MP174865IS or type

MP174865 Lithium ion rechargeable cells all encapsulated in Wacker Elastosil RT675.

17.4

The products covered by this certificate incorporate previously certified devices, it is therefore the

responsibility of the manufacturer to continually monitor the status of the certification associated with

these devices, and the manufacturer shall inform Sira of any modifications of the devices that may

impinge upon the explosion safety design of their products.