Specifications – ReliOn HEM-741CREL Manuel d'utilisation
Page 19

19
SPECIFICATIONS
Model:
ReliOn HEM-741CREL
Display:
LCD Digital Display
Measurement Range:
Pressure: 0 to 280 mmHg
Pulse: 40-180/min.
Accuracy/Calibration:
Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading
Inflation:
Automatic by electric pump
Deflation:
Automatic pressure release valve
Rapid Air Release:
Automatic exhaust valve
Pressure Detection:
Capacitive pressure sensor
Measurement Method:
Oscillometric method
Pulse Wave Detection:
Capacitance type pressure sensor
Power Source:
4 "AA" batteries
AC Adapter
(sold separately model HEM-ADPT1)
Battery Life:
Approx. 300 uses when used once
a day for 2 minutes
Operating Temperature/
50°F to 104°F (10°C to 40°C)
Humidity:
30 to 85% RH maximum
Storage Temperature/
-4°F to 140°F (-20°C to 60°C)
Humidity:
10 to 95% RH maximum
Console Weight:
Approximately 13 oz. (350g) with cuff,
excluding batteries
Outer Dimensions:
Approx. 5" (l) x 4 3/4" (w) x 3 3/4" (h)
(125mm x 119mm x 94mm)
Cuff Dimensions:
Approx. 5 1/2" (w) x 18 7/8" (l)
(140mm x 480mm)
Arm Circumference Range:
Approximately 9'' to 13'' (220mm - 320mm)
Contents:
Monitor, Arm cuff,and Instruction Manual
UPC Code:
0 73796 61115 6
NOTE: These specifications, to improve performance, are subject to change without notice.