Chapter 5: maintaining your waterpro ro system – Labconco WaterPr RO Systems 90750 Series Manuel d'utilisation

Page 33

Advertising
background image

Chapter 5: Maintaining Your WaterPro RO System

29

Original Instructions


Front Cover Removal

1. Disconnect the power cord.
2. Remove the two black screws on the front panel.
3. Raise the front of the cover slightly and pull out.


Prefilter and Carbon/Prefilter Replacement

NOTE: Prefilter – change at least once a month, carbon/prefilter change at least once
every two months.

1. Shut off feed-water supply and release pressure in the system by pressing the dispense

valve button.

2. With the Spanner wrench provided, remove the first bowl on the right to replace the

Prefilter and the middle bowl to replace the Carbon filter.

3. Remove the old filter.
4. Discard water in bowl.
5. Clean the bowl with mild soap and water, rinse and towel dry.
6. Insert new filter and place bowl on unit.
7. Hand tighten bowl in place. Do not use wrench to tighten as this may damage the bowl.


Air Vent Filter Replacement

Remove old filter and discard. Press new filter in tubing.

Advertising