Caractéristiques techniques, Ftb-635, Copper specifications – EXFO FTB-635 Wideband Copper and DSL Test Module Manuel d'utilisation

Page 400: Continued)

Advertising
background image

Caractéristiques techniques

392

FTB-635

COPPER SPECIFICATIONS

a, b, c

(continued)

Isolation resistance (stress/leakage) (continued)

Source

50 V to 500 V (current safely limited to 0.5 mA)

Soak timer

1 s to 59.9 min

VF noise measurement

Frequency range

200 Hz to 20 kHz

Level range

–90 dBm to 20 dBm

Resolution

0.1 dB

Uncertainty (accuracy)

–90 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 20 dBm, uncertainty (accuracy) ±1 dB

Filters

ITU: none, psophometric, P-notched, 3.4 kHz, D-filter, 15 kHz
ANSI: none, C-message, C-notched, 3.4 kHz, D-filter, 15 kHz

Impedance

600

1

VF impulse noise

Low threshold

–40 dBm to 0 dBm, in 1 dB steps

Mid threshold

Low threshold plus separation

High threshold

Mid threshold plus separation

Test duration

Minutes: 1, 5, 10, 15, 30, 60
Hours: 4, 8, 12, 24, 100

Separation

1 dB to 6 dB, in 1 dB steps

Dead time

125 ms

Filters

None, 3 kHz flat, C-message, psophometric,
notched and D-filter (IEEE 743-1995)

Counter

Maximum 999 for each threshold

Timer

1 min to 24 h, default is 15 min

Power influence (noise to ground)

Noise range

–60 dBm to 10 dBm

Uncertainty (accuracy)

–60 dBm to –50 dBm ± 2 dB
–50 dBm to 10 dBm ± 1 dB

Graphical display

Third triplet harmonics to 20 kHz

VF longitudinal balance

Frequency

1004 Hz

Level range

0 dB to 100 dB

Level uncertainty (accuracy)

±1 dB

Impedance

600

1

Time-domain reflectometer (TDR)

Modes

Automatic, Manual, and Crosstalk (Xtalk) operation with location of most significant event(s)

Distance range

0 m to 6700 m (0 ft up to 22 000 ft) on 24 AWG (0.5 mm) cable

Pulse width

15 ns to 20 μs (automatic control)

Amplitude

7.5 V p-p on cable, 9 V p-p open circuit

Velocity of propagation (VOP)

0.40 to 0.99

Distance uncertainty (accuracy)

e

±(0.5 m + 1 % x distance)

Units

Meters and feet

NOTES

a. Subject to change without notice.

b. Typical, at 23 °C ± 3 °C, on batteries, with no USB connection.

c. Specifications based on 24 AWG (PE 0.5 mm) cabling.

d. From 10 mA to 110 mA.
e. Qualified up to 300 m (1000 ft) and does not include the uncertainty due to VOP.

Measurement

Range

Resolution

Uncertainty (accuracy)

DC voltage

0 to 400 V

0.1 V for 0 to 99.9 V
1 V for 100 to 400 V

±(|1 %| + 0.5 VDC)

AC voltage

0 to 280 Vrms

0.1 VAC for 0 to 99.9 VAC
1 VAC for 100 to 280 VAC

±(1 % + 0.5 VAC)

Isolation resistance
(stress/leakage)

0 to 1 G

1, auto-ranging

1 k

1 to 99 M1

100 M

1 to 999 M1

Three digits

±(2 % + 1 digit)
±(5 % + 1 digit)

Resistance

0 to 100 M

1

0 to 999

1

1 k

1 to 100 M1

Three digits

±(1 % + 5

1)

±(2 % + 1 digit)

Capacitance

0 nF to 2 μF

Four digits

±(2 % + 50 pF)

DC current

0 to 110 mA

0.1 mA

±(|2 %| + 1 mA)

AC current

0 to 110 mA

0.1 mA

±(|2 %|+ 1 mA)

d

Station Ground

0 to 1 M

1

0 to 999

1

1 k

1 to 1 M1

Up to three digits

±(1 % + 3

1)

±(2 % + 1 digit)

Advertising