Parts List

Figure 1

OPERATING INSTRUCTIONS

1.

Connect instrument head to the handle.

2.

Depress On/Off Button on Rheostat Section and rotate the Light Intensity Control
Ring clockwise (CW). To intensify light, continue to rotate until stop is reached
(Figure 1).

3.

To turn off, rotate counter clockwise (CCW). Button will click when fully off. Turn
fully off after each use to ensure maximum on-time from the battery.

CHARGING INSTRUCTIONS

7114x Universal Charger

1.

Insert the handle into the Well. Charge for 24 continuous hours to charge a dead
battery (less to top off).

2.

Fully charged handles may remain in the Charger if desired.

3.

Charging the Battery after partial discharge does not negatively affect overall
battery memory (unlike Ni-Cad cells).

71110 Desk Charger (obsolete) and Ophthalmic Stands

See well adapter instructions for installing well adapter(s). Once adapters are installed,
follow charging procedures indicated under the 7114x Universal Charger.

NOTE: The Battery Section is not shipped fully charged. For maximum performance
charge the Battery Section for 24 hours when new or after a long period of inactivity.

LOW BATTERY INDICATOR

1.

The Low Battery Indicator will illuminate several minutes before handle shuts off
depending on the age of the battery (Figure 1).

2.

Once the battery is fully discharged, the Low Battery Indicator may flicker.

INSTALLING 71980 WELL ADAPTERS INTO THE 71110
DESK CHARGER

1.

Insert the Well Adapter into one of the Handle Receptacles in the Desk Charger.
Be sure to insert the Adapter until it is fully seated. Once the Well Adapter has
been installed, it does not need to be removed (Figure 2).

2.

One or two Well Adapters may be used with the Desk Charger.

BATTERY REPLACEMENT

Removing the old Battery Section:

1.

Hold the instrument handle firmly by the Battery Section.

2.

Turn the Rheostat Section Counterclockwise until it detaches from the Battery
Section (see figure 2).

Installing A New Battery Section

1.

Insert the Rheostat Section into the new Battery Section.

2.

To connect the two sections, turn the Rheostat Section clockwise until tight (see
figure 2).

3.

Verify the Rheostat Section is secure on the Battery Section of the instrument
handle.

CLEANING

The cleaning of all Welch Allyn power handles and chargers is easily accomplished by
wiping the external surface with a cloth dampened with a mild detergent and water
solution, 70% isopropyl alcohol, or a 10% bleach solution (by volume). Do not immerse.

NOTE: Excess solution entering the assembly could damage internal components. Use
caution to ensure cloth is not saturated with solution.

Do not sterilize the 719 Series Power Handle or Well Adapter.

Battery Recycling

Welch Allyn now employs the services of an agency which can disassemble and recycle
all components of rechargeable batteries so that nothing gets landfilled or incinerated.
Should you wish to dispose of your expended Welch Allyn rechargeable battery via
recycling (in the U.S. and Canada only), please send to:

Welch Allyn

Welch Allyn Canada

RECYCLE BATTERY

RECYCLE BATTERY

4341 State Street Road

160 Matheson Blvd. East

Skaneateles Falls, NY

Mississauga, Ontario

13153-0220

CANADA L4Z 1V4

For locations outside the U.S. and Canada consult your local recycling authority for
correct disposal of batteries.

SPECIFICATIONS

(Specifications and design may change without notice)

Handle

Model Number:

71911

Dimensions:

5.4” Long x 1.125” Diameter (Max.)(137mm x 28.6 mm)

Weight:

4.5 oz. (128 g)

Typical Charge Time:

24 hours (fully discharged)

Typical Discharge Time:

110 minutes (for new battery)

Approvals:

Complies with: IEC 60601-1, CAN/CSA C22.2 No 125 and

UL544

When used with 71980 Well Adapter and 711 series
Desk Charger

Equipment Classification:

TYPE BF Equipment.

Equipment Classification

(All Charger Model Numbers)

Double Insulation

Equipment Classification

(All System Components)

IXPO

Equipment not protected against the ingress of water.

Operating Mode:

Continuous Operation

NOTE: This product complies with current required standards for electromagnetic
interference and should not present problems to other equipment or is affected by other
devices. As a precaution, avoid using this device in close proximity to other equipment.

Environment

(

All System Components

)

Transport/Storage:

-20°C ~ 49°C, 95% R.H. Max, 500hPa - 1060hPa, Altitude

Operating:

10°C ~ 40°C, 95% R.H. Max, 500hPa - 1060hPa, Altitude

The CE mark on this product indicates it has been tested to and
conforms with the provisions noted in the 93/42/EEC Medical Device
Directive.

Authorized European Representative:
European Regulatory Manager
Welch Allyn, Ltd.
Navan Business Park
Dublin Road
Navan, County Meath, Republic of Ireland
Telephone: 353 46 67700
Fax: 353 46 27128

SERVICE AND WARRANTY
Service Information

Other than battery replacement, this equipment contains no user servicable parts. Return
to a Welch Allyn authorized service center for service.

For Technical Support or to obtain return instructions, please contact your nearest Welch
Allyn service center listed below:

Welch Allyn, Inc.

Welch Allyn, Ltd - Canada

4341 State Street Road

160 Matheson Blvd. E., Unit #2

Skaneateles Falls, NY 13153-0220

Mississauga, Canada L4Z 1V4

Telephone: 800-535-6663

Telephone: 905-890-0004

Fax: 315-685-4653

Fax: 905-890-0008

Welch Allyn, Ltd. - UK

Welch Allyn, GmbH PSC

Cublington Road

Zollerstrasse 2-4

Aston Abbotts

D72417 Juningen, Germany

Buckinghamshire England HP22 4ND

Telephone: 011-49-7477-927186

Telephone: 011-0207-365-6780

Fax: 011-49-7477-927293

Fax: 011-0207-365-9694

Welch Allyn, LTD - Singapore

Welch Allyn China Service Center

300 Beach Road

Room 708 Central Plaza

The Concourse #25-08

No. 277 Huang Pi Bei Rd.

Singapore 199555

Huang Pu District

Telephone: 011-656-291-0882

Shanghai 20003, China

Fax: 011-656-291-5780

Telephone: 011-86-21-63279631
FAX: 011-86-21-63279632

Welch Allyn, Ltd. - Australia
The Metro Centre Unit 5
38-46 South Street
Rydalmere NSW 2116, Australia
Telephone: 011-(612) 9638-3000
Fax: 011-(612) 9638-3500

Warranty

The 719 series Power Handle is guaranteed by Welch Allyn against all manufacturing
defects. Welch Allyn will repair or replace, free of charge, any parts of its own
manufacture proven to be defective through causes other than misuse, neglect, damage
in shipment, or normal wear.

Welch Allyn warrants that the Welch Allyn 71960 Battery Section will perform to original
specifications during the two years from the date of manufacture, when used with the
Welch Allyn 719 series Power Handle. This warranty does not apply to the use of the
71960 battery in other products. A defective battery will be replaced if it fails within two
years of the date of manufacture.

On/Off
Button

Light
Intensity
Control
Ring

Low
Battery
Indicator

Rheostat
Section

Battery
Section
71960

Chrome Ring

Instrument
Head
Attachment

Figure 2

US

Advertising