Product description

Centrifuge 5430/5430 R

English (EN)

17

*) Centrifugation of round-bottom tubes and blood taking systems can be done using additional adapters
(see p. 130).

To determine the g-force (rcf) for a special adapter, you can perform a calculation according to DIN 58 970
with the following formula:
rcf = 1.118 · 10

-5

· n

2

· r

max

n: speed in min

-1

(rpm)

r

max

: max. centrifuging radius in cm

Example:

In Rotor FA-45-30-11, the 0.5 mL adapter has a maximum radius of 8.4 cm. At 7,000 rpm, a maximum
g-force of 4,600 × g is achieved.

Rotor F-35-6-30*

For 15 mL conical tubes

11.0

7,540

For 50 mL conical tubes

10.5

7,197

for Centriplus centrifuge filter units

11.1

7,567

A-2-MTP rotor

Without adapter

9.0

2,204

for 384 well PCR plates

7.7

1,885

for 96 well PCR plates

7.3

1,788

CombiSlide adapter

7.7

1,885

Rotor

Adapter

Max.centrifugation
radius r

max

[cm]

Max. g-force (rcf)

Advertising