6 closing the centrifuge lid, Closing the centrifuge lid – Eppendorf 5430 R Centrifuge Manuel d'utilisation

Page 35

Advertising
background image

Operation

Centrifuge 5430/5430 R

English (EN)

35

6.

Load the buckets symmetrically.

Abb. 5-3: Incomplete, but symmetric loading of the buckets. The pegs of each bucket must be uniformly loaded.

Fig. 5-3:

Incomplete, but symmetric loading of the buckets. The pegs of each bucket must be uniformly
loaded.

Abb. 5-4: Symmetrical loading of the plates.

Fig. 5-4:

Symmetrical loading of the plates.

The plate arrangement shown on the right-hand side is incorrect, as the buckets will not swing properly.

The plates have some play in the buckets.

7.

Check the loading of the bucket.

5.4.6

Closing the centrifuge lid

1.

Check the correct attachment of the rotor and rotor lid.

2.

Push down the centrifuge lid until the lid latch engages and the lid is automatically closed.

The centrifuge will close automatically.

The

open key lights up blue. The symbol appears in the display.

 Carry out a brief centrifugation test at low speed (e.g., 1000 rpm) when you use a tube or

plate type for the first time.

Only 5430 R: When using the A-2-MTPs, centrifuge without the upper shell of the wind shield
in order to guarantee precise and quick temperature control of samples. Note that the
centrifugation noise will increase slightly in this case.
This does not apply to the S-24-11-AT. The S-24-11-AT must always be operated with a rotor
lid.

WARNING! Risk of injury when opening or closing the centrifuge lid.
There is a risk of crushing your fingers when opening or closing the centrifuge lid.
 When opening or closing the centrifuge lid, do not reach between the lid and device or into

the latching mechanism of the lid.

 Always open the centrifuge lid completely to prevent it from falling.

Advertising
Ce manuel est liée aux produits suivants: