Centrifuge 5418 / 5418 R

English (EN)

17

4

Installation

4.1

Selecting the location

Select the location for the device according to the following criteria:

• Suitable power connection as per the name plate (230

V/120

V/100

V).

• Stable, horizontal and resonance-free lab bench. Weight of the device: 7.7 kg (5418) or 22 kg (5418 R).
• A well ventilated environment which is protected from direct sunlight to prevent the device from heating

up more.

4.2

Preparing installation

Prerequisites

The weight of the Centrifuge 5418 is 7.7

kg or the Centrifuge 5418 R 22

kg. For unpacking and installing

the Centrifuge 5418 R, you require the assistance of another person.

Perform the following steps in the sequence described.

1. Open the packaging board.

2. Remove the accessories.

3. Lift the Centrifuge 5418 / 5418

R out of the carton. Lift the Centrifuge 5418 R out of the carton with the

help of another person.

4. Remove the front and back transport protection pads.

5. Place the device on a suitable lab bench.

6. Remove the plastic sleeve.

NOTICE! If an error occurs, the objects in the immediate proximity of the device will be
damaged.

In accordance with recommendations in EN 61010-2-020, leave a safety clearance of

30

cm around the device during operation.

Please remove all materials and objects from this area.

NOTICE! Damage from overheating.

Do not install the device near to any heat sources (e.g., heating, drying cabinet).

Do not expose the device to direct sunlight.

Ensure unobstructed air circulation. Keep free a clearance of at least 33

cm around all

ventilation grilles.