61

11 Annexe

VEGAPULS 62 version tube guide d'ondes • Protocole Modbus et Levelmaster

41363-FR-121120

Response:

Paramètres

Length

Code/Data

Set Receive to Transmit Delay

5 characters ASCII

UuuFn
n = number of measurement values

(0, 1 or 2)

Report Receive to Transmit Delay

Request:

Paramètres

Length

Code/Data

Report Receive to Transmit Delay

4 characters ASCII

UuuR

Response:

Paramètres

Length

Code/Data

Report Receive to Transmit Delay

7 characters ASCII

UuuRmmm
mmm = milliseconds (50 bis 250),

default = 127 ms

Codes de défaut

Error Code

Name

EE-Error

Error While Storing Data in EEPROM

FR-Error

Erorr in Frame (to short, to long, wrong data)

LV-Error

Value out of limits