Specifications (continued) – Sony UP-2100 Manuel d'utilisation

Page 76

Advertising
background image

76

Others

Note

When AUTO LIVE of the FUNCTION-SETUP
menu is set to OFF, the memory image is
displayed on the screen after the image is captured
in memory. In such a case, whenever you store the
image, press the SOURCE/MEMORY button so
that the source image appears on the video
monitor. Displaying the memory image does not
allow to capture the image with the correct timing.

When IMMED. CAP. is set to ON

Note

This mode is effective only the source image is
displayed on the monitor screen. If the memory
image is displayed, the image will not be captured
even if the pulse is sent to the printer. Be sure to
display the source image when sending the pulse.

Design and specifications are subject to change
without notice.

V-SYNC

0

1

3

8

2

t

RPE

t

RPW

v

RP

Notation

tRPE

tRPW

VRP

Parameter

Time within which
a remote control
pulse should be
cleared to zero.

Length of a
remote control
pulse.

Amplitude of a
remote control
pulse.

TYP

MIN

3

100

Remarks

This value
is based
on TTL
standards.

Unit

msec

msec

V

MAX

10

5

Captured image

Remote
control
pulse
signal

Specifications (continued)

Using the automatic printing capabilities
(REMOTE 2)

If you send the remote control pulse signals
illustrated through the REMOTE 2 connector, the
printer is remotely controlled according to the
remote control setting. (see page 65)
Turn on the power of the printer and display the
source image on the monitor screen. Send a remote
control pulse signal at the timing shown below.
The timing depends on the setting on item
IMMED. CAP of the FUNCTION SETUP menu.

Remote 2 connector pin assignment

GND

BUSY

MIN

3

2

1

CN5

FL1

FL2

Note

The remote control pulse signal examples
introduced here are one of the typical operation
timing. The timing may be affected due to the
selected memory page and print type.

When IMMED. CAP. is set to OFF

V-SYNC

0

1

5

6

7

8

9

Captured image

Notation

tRPE

tRPW

VRP

Parameter

Time within which
a remote control
pulse should be
cleared to zero.

Length of a
remote control
pulse.

Amplitude of a
remote control
pulse.

TYP

MIN

3

100

Remarks

This value
is based
on TTL
standards.

Unit

msec

msec

V

MAX

12

5

MIN
(IN)

BUSY
(OUT)

Remote

control

pulse

signal

VRP

tRPE

tRPW

Advertising